Choline Assay Kit, BioAssay™

Cat# C5057-40-96T

Size : 96Tests

Marca : US Biological

Contatta il distributore locale :


Telefono : +1 850 650 7790


Shipping Temp
Dry Ice

Storage Temp
-20°C

Choline and its metabolites play important roles in membrane structure integrity, cellular signaling and cholinergic neurotransmission. Aberrant regulation in choline metabolism has been associated with mental illness such as anxiety. This method provides a simple, direct and high-throughput assay for measuring choline in biological samples. In this assay, free choline is oxidized by choline oxidase to betaine and H2O2 which reacts with a specific dye to form a pink colored product. The color intensity at 570nm or fluorescence intensity (530/585 nm) is directly proportional to the choline concentration in the sample.

Key Features:
Use 20ul samples. Linear detection range: Colorimetric Assay: 1 to 100uM, fluorometric assay 0.2 to 10uM choline.

Applications:
Assays: choline in biological samples such as serum, plasma, urine, saliva, milk, tissue, and cell culture.
Drug Discovery/Pharmacology: effects of drugs on choline metabolism.

Kit Components:
C5057-40A: Assay Buffer: 1x10ml
C5057-40B: Enzyme Mix: 1x1 vial
C5057-40C: Dye Reagent: 1x120ul
C5057-40D: Standard (2mM Choline): 1x400ul

Storage and Stability:
Store other components at -20°C. Stable for 6 months after receipt. For maximum recovery of product, centrifuge the original vial after thawing and prior to removing the cap.

Colorimetric Assay:
Sample treatment: liquid samples such as serum and plasma can be assayed directly. Tissue and cell lysates can be prepared by homogenization in cold 1 x PBS and centrifugation (5 min at 14,000rpm). Use clear supernatants for assay. Milk samples should be cleared by mixing 600ul milk with 100ul 6 N HCl. Centrifuge 5 min at 14,000 rpm. Transfer 300ul supernatant into a clean tube and neutralize with 50ul 6 N NaOH. The neutralized supernatant is ready for assay (dilution factor n=1.36).
Note: (1). SH-containing reagents (e.g. b–mercaptoethanol, dithiothreitol, > 5uM) are known to interfere in this assay and should be avoided in sample preparation. (2). This assay is based on an enzyme-catalyzed kinetic reaction. Addition of Working Reagent should be quick and mixing should be brief but thorough.
1. Equilibrate all components to room temperature. Briefly centrifuge the tubes before opening. Keep thawed tubes on ice during assay.
2. Standards: mix 12ul 2mM Standard with 228ul dH2O (final 100uM). Dilute standard in dH2O as follows.
No 100uM STD+H2O Vol (ul) Choline (uM)
1 100ul+0ul 100 100
2 60ul+40ul 100 60
3 30ul+70ul 100 30
4 0ul +100ul 100 0
Transfer 20ul diluted standards into separate wells of a clear flatbottom 96-well plate.
Samples: transfer 20ul of each sample into separate wells of the plate.
3. Color reaction. Prepare enough Working Reagent by mixing, for each reaction well, 85ul Assay Buffer, 1ul Enzyme Mix and 1ul Dye Reagent. Add 80ul Working Reagent to each well. Tap plate to mix. Incubate 20 min at room temperature.
4. Read optical density at 570nm (550-585nm).

Fluorometric Assay:
The Fluorometric Assay: is 10 times more sensitive than the colorimetric method. The procedure is similar to that for the Colorimetric Assay: except that (1) 0, 3, 6 and 10uM choline standards and (2) a black 96-well plate are used. Read fluorescence intensity at lex=530 nm and l em=585 nm.
Note: if the calculated choline concentration of a sample is higher than 100uM in the Colorimetric Assay: or 10uM in the Fluorometric Assay:, dilute sample in water and repeat the assay. Multiply result by the dilution factor n.

Calculation:
Subtract blank value (#4) from the standard values and plot the DOD or DF against standard concentrations. Determine the slope and calculate the choline concentration of Sample, [Choline] =
RSAMPLE–RBLANK
Slope (uM-1)
× n (uM)
RSAMPLE and RBLANK are optical density or fluorescence intensity
readings of the Sample and H2O Blank, respectively. n is the sample
dilution factor.
Conversions: 1mM choline equals 10.4 mg/dL, 0.010% or 104 ppm.

Materials Required, But Not Provided:
Pipetting devices, centrifuge tubes, clear flat-bottom uncoated 96-well plates, optical density plate reader; black flat-bottom uncoated 96-well plates, fluorescence plate reader.

Important Note
This product as supplied is intended for research use only, not for use in human, therapeutic or diagnostic applications without the expressed written authorization of United States Biological.