Mouse ACTH (Adrenocorticotropic Hormone) ELISA Kit

Referência E-EL-M0079-96T

Tamanho : 96T

Marca : Elabscience


Mouse ACTH(Adrenocorticotropic Hormone) ELISA Kit (E-EL-M0079)

loading Resources
Mouse ACTH(Adrenocorticotropic Hormone) ELISA Kit - 1
  • Mouse ACTH(Adrenocorticotropic Hormone) ELISA Kit - 1
  • Mouse ACTH(Adrenocorticotropic Hormone) ELISA Kit - 2
  • Mouse ACTH(Adrenocorticotropic Hormone) ELISA Kit - 3
All Size Price Qty
96T $ 495.00
- +
48T $ 396.00
- +
24T $ 150.00
- +
96T*5 Inquire /
96T*10 Inquire /

For research use only.

Product Summary
Sensitivity 9.38 pg/mL
Detection Range 15.63-1000 pg/mL
Sample Volume 50 μL
Total Assay Time 2 h 30 min
Reactivity Mouse
Specificity This kit recognizes Mouse ACTH in samples.No significant cross-reactivity or interference between Mouse ACTH and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Colorimetric method, ELISA, Competitive
Assay Type Competitive-ELISA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃
Expiration Date 12 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
1000 0.201 0.207 0.204 -
500 0.312 0.306 0.309 -
250 0.504 0.506 0.505 -
125 0.787 0.807 0.797 -
62.5 1.146 1.196 1.171 -
31.25 1.555 1.533 1.544 -
15.63 1.864 1.830 1.847 -
0 2.259 2.359 2.309 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 47.410 159.390 354.720 42.880 156.150 382.650
Standard deviation 2.170 8.130 24.830 3.300 10.880 29.620
CV(%) 4.570 5.100 7.000 7.700 6.970 7.740
Recovery
The recovery of Mouse ACTH was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 89-102 94
EDTA plasma (n=8) 88-104 95
Cell culture media (n=8) 89-102 96
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Mouse ACTH and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 90-103 96-112 89-103
Average 96 104 97
1:4 Range 100-116 88-100 85-97
Average 106 93 91
1:8 Range 99-113 88-98 85-100
Average 106 93 91
Stability
Each kit batch is subjected to accelerated stability testing and real-time stability monitoring. Sample performance is evaluated after storage at 37 °C for 10 days to assess the impact of elevated temperature on assay reliability and reagent integrity.
/ Variation range of 37°C mean
concentration / 2-8°C mean
Sample 1(n=16) 100.74-106.95
Sample 2(n=16) 87.79-94.56
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Mouse ACTH. During the reaction, Mouse ACTH in the sample or standard competes with a fixed amount of Mouse ACTH on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Mouse ACTH. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Mouse ACTH in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Gene ID 18976
Uniport ID P01193
Research Area Cancer , Metabolism , Neuroscience , Signaltransduction
Mouse ACTH(Adrenocorticotropic Hormone) ELISA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

Species

{{relatedElisaPage.numTotal}} Results

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

Videos

Competitive ELISA Operation Guide Video Sandwich ELISA Operation Guide Video
确定

Você também pode estar interessado nos seguintes produtos:



Referência
Descrição
Cond.
Price Bef. VAT